SSM 6000 trace gas analyser for measuring hydrogen sulphide in the ppb range: to protect adjacent facility components (e.g. catalytic converters) from H2S.
• highly sensitive analysis system with a cycle time of approximately two to five minutes (depending on the application)
• optionally available: additional oxygen trace gas analyser with a measuring range of 10 ppm O2
• complete with installed pipes on mounting plate
TECHNICAL DATA
Mounting plate dimensions (WxHxD): 900 x 500 x 350 mm
Protection: IP20
Measuring range: H2S: 0 to 2.000 ppb
Detection limit: H2S: 5 ppb
Resolution: H2S: 1 ppb
Display: LED display for measuring values and LC display
Communication: RS232, optionally Profibus DP
Operating temperature: +5 to +40 ° C
Power supply: 85 to 264 VAC, 47 to 63 Hz