branding

Biogas analysis technology

For biogas analysis, we offer solutions for biogas plants, sewage and landfill sites, laboratory facilities and custom applications

SSM 6000 LAB

The SSM 6000 was developed specifically for the analysis of biogenic process gases, such as biogas, sewage gas and landfill gas. It is designed for on-site plant use, capable of meeting regular process control requirements and unites the high-quality sensor technology used for continual gas analysis with multi-stage gas conditioning technology. If used to full capacity, all main gas types can be recorded: methane, hydrogen sulphide, oxygen, carbon dioxide and hydrogen. Analysis is fully automatic, appearing at variable time intervals (at 8-hour intervals, for example). Continuous measuring of individual types of gases is also optionally possible. The SSM 6000 is easy to operate, equipped with a clear display and a data logger with history function where all measured data is stored together with the precise recording time.

SSM 6000 LAB

A special version of the SSM 6000 that has been designed specifically for labor applications.

The SSM 6000 LAB device is specialised for measuring smallest sample volumes.


proCAL and automatic calibration

The proCAL method developed by PRONOVA enables even greater long-term stability for methane and carbon dioxide measurements. Thanks to additional automatic one-point calibration of all other measuring channels, the SSM 6000 is setting new standards in long-term stability.

ARTICLE AND PRIZES
Art.-Nr.
Article
Cart
1130 SSM 6000 LAB Request Price
1103 SSM 6000 Classic (CH4, H2S, O2) Request Price
1104 SSM 6000 Classic (CH4, H2S, O2, CO2) Request Price
Show Cart
TECHNICAL DATA
Measuring ranges:  
CH4: 0 to 100 Vol%
CO2:

0 to 100 Vol%

O2:

0 to 25 Vol%

H2S:

0 to 5,000 ppm SSM Classic

H2S:

0 to 1,000 ppm SSM LT

H2: 0 to 1,000 ppm
  others on request
Resolution:

 

CH4:

0.1 Vol%

CO2:

0.1 Vol%

O2:

0.1 Vol%

H2S: 1 ppm
H2: 1 ppm
Display: LED display for measuring values and LC display
Communication: RS232, Profibus DP option
Operating temperature:

+5 to +40 °C

Power supply:

85 to 264 VAC, 47 to 63 Hz

Protection class:

IP 20

Dimensions and weight:

300 x 400 x 200 mm,
    approx. 13 kg (Classic 4K)