branding

iRAS® – Water analysis technology

Our iRAS® products provide solutions for wastewater treatment plants, environmental protection, drinking water, ground and surface waters, laboratory, energy production, soil extracts, and much more

Gas sensitive electrodes-Carbondioxide

Carbondioxide-electrode

For the determination of dissolved oxygen according to the Clark-principle and carbon dioxid according to the Severinghouse-principle in water irrespective of colour or turbidity.

 

The carbondioxide-electrode with an integrated temperature sensor measures the dissolved carbondioxide in water and operates on the principle of Severinghausen. The electrode is a glas electrode provided with a CO2-permeable membrane. Between the platinum electrode and the reference. Because of the diffusion of CO2 through the membrane the pH value is changing and indicates the CO2 concentration.

 

Applications:

 

Food industry, Water management, Environment analytic, Biotechnology

ARTICLE AND PRIZES
Art.-Nr.
Article
Cart
40186005 Carbondioxide-electrode € 1547,00
0 bis 3.000, PG 16
Art.-Nr. Article Accessories Cart
40xxx100 1 m sealed cable without connector Request Price
40xxx101 1 m sealed cable with BNC connector (plug) Request Price
40xxx005 PG 13.5“ industrial connector Request Price
40xxx006 S7 lab connector Request Price
Show Cart
TECHNICAL DATA
Measurement range:

0 to 3,000 mg/l

Measuring principle:

Severinghaus

Precision:

 ± 4 % measured data

Setting time (20 °C):

< 4 min (90% des Messwertes)

Setting time (10 °C):   

< 8 min (90 % des Messwertes)

Interference ions:

SO32−, S2−

Temperature:

0 to +50° C

Membrane:

Teflon

Operating electrode:

Glass electrode with membrane

Reference electrode:

Ag/AgCl

Temperature sensor:

NTC 10 kΩ at 25 °C

Shaft material:

plastics (black), stainless steel

Dimension (length x ø):

173 mm x 18 mm
    immersion depth: 120 mm

product

PET 2000

The PET 2000 activity meter can measure the potential absorption of nutrient salts under the same conditions as the actual nutrient salt absorption through the roots. This means that the dissolved salt content is measured in the soil and... ...more Topseller

Consumers are shown prices including VAT (gross), and companies are shown prices excluding VAT (net).